HPLC Method

  • High Performance Liquid Chromatography (HPLC) is considered as the “Golden Standard” method for the HbA1c test
  • Ensure that the test results aligned to the Diabetes Control and Complications Trial (DCCT) and the UK Prospective Diabetes Study (UKPDS) reference studies

Accurate

  • CV ≤ 1%, provides better repeatability
  • NGSP and IFCC certificate
  • Ability to separate common variants, HbE, HbS, HbC, and HbD, providing more information for clinical decisions

Efficient

  • Only 45 seconds for fast mode
  • Automatic barcode reading
  • Automatic mixing
  • Automatic cap-piercing
  • Automatic printing
  • Diluted and Whole Blood Sample Auto-switching

LIS specifications

  • Graph can be transmitted to the LIS system for easy query
  • Supporting bidirectional transmission of results
  • Supporting multi-machine connection, and providing higher productivity for large laboratories

Easy to use

  • Fully automatic, no manual operation required after running samples
  • The 12.1-inch touch screen provides a clear interactive interface to avoid operator error

Principle

Ion Exchange HPLC

Analytes

HbA1c, HbF, HbE, HbD, HbS, HbC, HbA1a, LA1c, HbA0

Speed

45 sec/test for fast mode
60 sec/test for normal mode
110 sec/test for variant mode

Sample Consumption

Whole blood: 4 μL

Loading Capacity

100 Samples

Measuring Range (NGSP)

3.0%–20.1%

Precision

CV < 1%

Data Storage

On-board memory up to 100,000 results with graph

On Board Reagent Volume Detection

Weight Level Sensor

Display

12.1 inch touch screen

Output

Internal Thermal Printer
LIS System
USB

Sample Handler

Dilution and Whole Blood Sample Auto-switching
Internal barcode reader
Auto-mixing
Cap Piercing

Working Environment

10°C–30°C
Humidity: ≤80%
Altitude: Self-adaptive

Dimension

536 mm × 580 mm × 538 mm (L*W*H)

Weight

40.3 kg